Electrochemical Flow Cell: How a Flow Cell Works, Design, and Applications
An electrochemical flow cell is a reactor where flowing electrolyte undergoes electrochemical reactions at electrode surfaces. This article covers cell design, electrode spacing, membranes, flow fields, and reference electrodes, and their influence on performance and applications in electrolysis.
Hydrogen fuel cells and electrochemical energy conversion
This article was first published on
www.wevolver.comKey Takeaways
An electrochemical flow cell pumps electrolyte through a narrow gap between closely spaced electrodes. Compared with a batch cell, this configuration increases the electrode surface-to-volume ratio and reduces ohmic drop, potentially lowering the need for supporting electrolyte.
Parallel-plate cells are the most common design in organic electrochemistry. Laboratory models typically use electrode areas of 10–100 cm², linear flow rates of 1–20 cm/s, and inter-electrode gaps of 1–5 mm.
Mass transport limits reaction rates. In one well-studied laboratory cell, a flat-plate electrode achieved a mass transfer coefficient near 10⁻³ cm/s. Turbulence promoters increased conversion rates by 1.5–3 times, while a 4 mm nickel foam electrode achieved more than 20 times the mass transfer rate.
Membranes and flow-field designs involve trade-offs. A redox flow battery study using carbon paper electrodes measured an area resistance of 0.5 Ω·cm² in a flow-by design compared with 3.5 Ω·cm² in a flow-through design. Zero-gap CO₂ electrolyzers can use anion-exchange, cation-exchange, or bipolar membranes.
The term "flow cell" also describes devices outside electrochemistry, including Illumina sequencing flow cells that carry DNA template clusters and optical flow-through cuvettes.
Introduction
An electrochemical flow cell is a reactor in which a pump moves electrolyte through a channel between closely spaced electrodes, rather than allowing the solution to remain stationary in a beaker. The narrow inter-electrode gap, often less than 1 mm in some designs, provides a high electrode surface area relative to the reaction volume. In single-pass operation, the electrolyte flows continuously through the reactor, allowing electrochemical reactions to occur as the solution passes the electrode surfaces.
The advantages of flow are practical. Conventional beaker and H-cell electrolysis can be limited by mass transport, reaction rates, selectivity, and reproducibility. Flow configurations can address some of these limitations by controlling electrolyte movement, improving reactant transport to electrode surfaces, and providing more consistent reaction conditions. Flow electrolysis is also established in industrial processes, including the chlor-alkali industry.
The growing use of flow-based electrochemical systems reflects the need for greater control over reaction conditions, mass transport, and electrode utilization. By integrating electrolyte circulation with carefully designed electrode assemblies, these reactors offer a practical approach to laboratory-scale electrolysis, process development, and continuous chemical production.
Recommended reading: Flow Battery: How It Works, Types, Cost, and Lifespan
What Is a Flow Cell in Electrochemistry?
In electrochemistry, a flow cell is a reactor in which electrolyte is continuously pumped through a channel or chamber between electrodes. An electrochemical cell consists of electrodes, an electrolyte, and an electrical circuit that controls or measures the current and potential during a reaction. In a flow configuration, the movement of electrolyte through the reactor becomes an integral part of how the electrochemical reaction is conducted.
The geometry is central to its operation. Compared with many batch cells, flow reactors can provide a higher electrode surface area relative to electrolyte volume and shorter inter-electrode distances. These features can improve mass transport and reduce ohmic losses, potentially lowering the amount of supporting electrolyte required. Batch cells, by comparison, often have larger electrode separations and less efficient mass transport, which can limit reaction rates and conversion.
Working, Counter, and Reference Electrodes
A basic electrochemical cell uses two electrodes: a working electrode and a counter electrode. The working electrode is where the reaction of interest takes place, while the counter electrode completes the electrical circuit by supporting the complementary reaction. In a two-electrode arrangement, the potential difference is measured or controlled across both electrodes, so the counter electrode's potential may influence the overall cell voltage.
For more precise potential control, a three-electrode arrangement separates these functions. A reference electrode provides a stable potential against which the working electrode is measured, while an auxiliary (counter) electrode carries the current needed to complete the circuit.
In this arrangement, the potentiostat controls or measures the working electrode potential relative to the reference electrode, while current flows between the working electrode and the counter electrode. If reduction occurs at the working electrode, oxidation generally occurs at the counter electrode, or vice versa. The counter electrode reaction must be selected and managed to prevent its products from interfering with the reaction of interest.
Electrochemical experiments generally operate under either potentiostatic or galvanostatic control. Under potentiostatic conditions, the working electrode potential is controlled, and the resulting current is measured. Under galvanostatic conditions, the current is controlled and the resulting potential is measured. The two quantities are not independently controlled in these conventional modes.
Recommended reading: Microbial Fuel Cell: How It Works, Efficiency, and Applications
Flow Cell vs. Beaker Cell and H-Cell
Beaker cells, pot cells, and H-cells remain widely used in academic electrochemistry, including organic electrosynthesis, because they are relatively simple to assemble and operate. However, conventional batch configurations can struggle to achieve high conversion rates when mass transport is limited or the electrode area is small relative to the electrolyte volume. Depending on the reaction and operating conditions, achieving high conversion may require several hours or even days.
Two factors help explain these limitations: the mass transfer coefficient ((k_m)), which describes how efficiently reactants reach the electrode surface, and the electrode area-to-volume ratio ((A/V)). Many laboratory batch cells have an (A/V) ratio well below 1 cm⁻¹, limiting the available electrode surface relative to the solution volume.
Flow configurations address these constraints through closely spaced electrodes, controlled electrolyte circulation, and reactor geometries designed to improve mass transport. The differences between common cell types are summarized below.
Property | Beaker cell or H-cell | Parallel-plate flow cell | Extended-channel or microflow cell |
Inter-electrode gap | Often relatively large | Gaps of 1–5 mm are readily achievable; some designs use gaps below 1 mm | Approximately 200 µm in the Syrris microflow cell and 0.5 mm in Ammonite cells, microfabricated chips can use gaps as narrow as 10 µm. |
Electrode area-to-volume ratio | Often below 1 cm⁻¹; concentric electrode arrangements can reach approximately 0.25 cm⁻¹ in some beaker configurations. | High electrode area relative to reaction volume | Extended channels maintain surface-to-volume ratios along the channel, although increasing channel length can raise pressure drop. |
Time for high conversion. | 95% conversion may take hours or days, depending on mass transport and electrode area. | A 10 × 10 cm cell could achieve 95% conversion of 1 L of 0.1 M reactant in under 2 hours under the conditions estimated in the cited study. | More than 95% conversion in a single pass at 1 mL/min was reported for the Syrris microflow cell |
Operating mode | Batch | Single-pass or recirculation | Often designed for high conversion in a single pass |
Main limitation | Often limited by mass transport and relatively low electrode area-to-volume ratio | Gas bubbles, precipitates, and poorly soluble species can complicate operation | Narrow channels can clog, particularly when solids form or precipitate |
How an Electrochemical Flow Cell Works
A laboratory electrochemical flow cell typically operates within a fluid circuit comprising a reservoir, pump, reactor, and outlet. Electrolytes move through the system, interact with the electrode surfaces, and exit as a product stream or return to the reservoir. Each stage influences the reaction conditions and overall performance.
Supply: Electrolyte is stored in an external reservoir and delivered to the reactor by a pump, often a peristaltic pump. In one CO₂ electrolysis study, catholyte and anolyte were circulated through the cell at 50 mL/min from 200 mL reservoirs fitted with cooling jackets.
Distribution: The electrolyte enters through an inlet channel, entry length, or flow distributor designed to spread the solution evenly across the electrode surface and minimize uneven flow.
Reaction: The electrolyte passes through the gap between the working and counter electrodes, where the electrochemical reaction occurs. In a well-designed parallel-plate cell, the working electrode surface is positioned uniformly relative to the counter electrode, helping maintain a consistent potential drop across the electrolyte. During single-pass operation, however, reactant concentration and local reaction rates can decrease along the channel.
Control: A potentiostat or power supply drives the reaction. Potentiostats provide precise potential control and are useful for characterization and small-scale electrolysis, although their current capacity may limit larger reactions. For many electrosynthesis applications, a constant-current power supply is sufficient, provided the selected current is appropriate for reactant concentration, electrode area, and mass transport.
Exit: In single-pass operation, the product stream leaves the reactor for collection or downstream processing. In recirculation mode, the electrolyte returns to the reservoir and passes through the cell again.
Single-Pass vs. Recirculation Operation
Recirculation allows electrolyte to pass through the cell repeatedly, often providing greater overall conversion without requiring a large reactor or reservoir flow rate. Higher circulation rates can also help remove gas bubbles from electrode surfaces. However, prolonged exposure to the electrodes may promote secondary reactions, product degradation, or unwanted byproduct formation, depending on the chemistry.
In industrial electrolysis, high flow rates are often used to improve mass transport, even when conversion per pass is relatively low. The electrolyte is then recycled between the reservoir and the reactor to achieve the desired overall conversion.
Single-pass operation enables continuous product collection and integration with downstream reactions, separation, or work-up processes. Its performance depends on balancing flow rate, current, and reactant concentration. Higher flow rates increase throughput but may reduce conversion per pass, while gas evolution can disrupt flow distribution and create localized resistance. As reactant is consumed along the channel, the local current density may also decrease from inlet to outlet.
Worked Example: Sizing the Current with Faraday's Law
Faraday's law relates the electrical charge passed through an electrochemical cell to the amount of substance converted. The Faraday constant (F) represents the charge carried by one mole of electrons and is approximately 96,485.33212 C/mol.
For a continuous flow reaction, the theoretical current required to convert a reactant stream can be estimated from its molar flow rate:
I = n × F × c × q
Where:
I = theoretical current (A)
n = number of electrons transferred per molecule of reactant
F = Faraday constant (96,485.33212 C/mol)
c = reactant concentration (mol/L)
q = volumetric flow rate (L/s)
Consider a laboratory microflow cell developed by Syrris, featuring an extended channel approximately 70 cm long and 1.5 mm wide, with an inter-electrode gap of approximately 200 µm. More than 95% conversion was reported at a flow rate of 1 mL/min. Using this flow rate as an illustrative calculation, assume a reactant concentration of 0.1 M and a two-electron oxidation.
Step 1: Convert the volumetric flow rate to L/s.
q = 1 mL/min ÷ 60,000
q = 1.67 × 10⁻⁵ L/s
Step 2: Calculate the molar flow rate.
Molar flow rate = c × q
= 0.1 mol/L × 1.67 × 10⁻⁵ L/s
= 1.67 × 10⁻⁶ mol/s
Step 3: Calculate the theoretical current.
I = n × F × c × q
= 2 × 96,485.33212 × 0.1 × 1.67 × 10⁻⁵
≈ 0.32 A
This estimate assumes 100% Faradaic efficiency, meaning all the charge contributes to the intended reaction. In practice, the required current may be higher if competing reactions reduce Faradaic efficiency.
Increasing the flow rate can improve throughput but may reduce conversion per pass. In the reported Syrris cell, conversion decreased to approximately 80% at 3 mL/min, while the reported production rate was 2 g/h. Current must therefore be matched to the flow rate, reactant concentration and active electrode area. Dividing the applied current by the active electrode area gives the average current density, although local current density may vary along the channel as reactant is consumed.
Checking the current against mass transport
The theoretical current must also be considered against the mass transfer limit. For a mass-transfer-limited reaction, the limiting current density can be estimated using:
iL = n × F × km × c
Where:
iL = limiting current density (A/cm²)
n = number of electrons transferred
F = Faraday constant (C/mol)
km = mass transfer coefficient (cm/s)
c = reactant concentration (mol/cm³)
For a flat-plate electrode with a mass transfer coefficient close to 10⁻³ cm/s, a reactant concentration of 0.1 M (equivalent to 10⁻⁴ mol/cm³) and a two-electron reaction:
iL = 2 × 96,485.33212 × 10⁻³ × 10⁻⁴
iL ≈ 0.0193 A/cm²
iL ≈ 19.3 mA/cm²
This estimate illustrates the mass transport limitation of a flat-plate electrode under the assumed conditions.
In a separate estimate, Pletcher's group considered a 10 × 10 cm parallel-plate electrolyzer processing 1 L of a 0.1 M reactant solution. An initial conversion rate of 0.2 mol/h could take the solution to 95% conversion in under two hours, assuming the reaction and mass transport conditions described.
For a two-electron reaction operating at 100% Faradaic efficiency, that initial rate corresponds to a theoretical current of approximately 10.7 A. Across a 100 cm² active electrode area, the average current density would be approximately 107 mA/cm².
This is substantially higher than the estimated flat-plate limiting current density of 19.3 mA/cm². Achieving such a reaction rate would therefore require enhanced mass transport, for example through turbulence promoters or a three-dimensional electrode structure, rather than relying on a flat plate under the same mass transfer conditions. The electrode area required for a reaction decreases as the sustainable current density increases, making mass transfer a key consideration when sizing an electrochemical reactor.
Electrochemical Flow Cell Design: Components and Architecture
Parallel-plate reactors are among the most widely used flow cell configurations in organic electrochemistry, and filter-press stacks are common in commercial electrosynthesis. A typical parallel-plate reactor consists of a spacer that defines the flow channel between the electrodes, insulating layers that separate the electrodes from the end plates, and a housing that compresses and secures the assembly.
Positioning the plates parallel to each other helps establish a more uniform current and potential distribution across the active area, which can improve selectivity and reproducibility. A simple design uses two electrode plates separated by an elastomer spacer with a cutout that forms the reaction chamber. Bolts compress the stack, while gaskets provide a seal against electrolyte leakage.
Component | Function | Examples from published cells |
End plates | Compress the assembly and accommodate fluid fittings | 10 mm aluminum plates fabricated using drilling and threading tools |
Insulating layers | Electrically isolate the electrodes from the end plates | Mylar foil |
Current collectors | Connect the electrodes to the power supply | 2.6 mm pogo pins, electrode tabs and titanium flow fields contacting gas diffusion electrodes |
Electrodes | Provide surfaces for the working and counter reactions | Plates, meshes, foams and felts made from graphite, carbon, nickel, stainless steel, platinum, glassy carbon or boron-doped diamond |
Spacer and gasket | Define the inter-electrode gap, form the channel and prevent leakage | Elastomer sheets, 0.3 mm laser-cut Mylar, PET and PTFE gaskets, including 13 µm gaskets used in detector cells |
Separates the anolyte and catholyte in divided configurations | Ion-permeable membranes, porous polymer separators and Nafion 117 | |
Flow field or distributor | Directs and distributes electrolyte across the active area | Inlet and outlet channels, flow distributors, serpentine and interdigitated channels |
Reference electrode port | Allows potential measurement and control in a three-electrode configuration | Ag/AgCl reference electrode connected through a Haber-Luggin capillary to the catholyte supply line |
The cell developed by Jud, Kappe and Cantillo illustrates how a reactor can be assembled with minimal machining. A stack of laser-cut Mylar foils functions as the gasket, insulation, inter-electrode spacer and flow channel, surrounding two 5 × 5 cm plate electrodes. Electrode thicknesses ranging from 0.1 to 6 mm can be accommodated by stacking 0.3 mm alignment gaskets. Ten stacked gaskets, for example, accommodate a 3 mm graphite plate.
Electrode Materials and Structure
Material selection should follow the reaction chemistry, operating conditions, and reactor configuration. Electrode material, surface area, porosity, and geometry all influence reaction selectivity, mass transport and current distribution.
In one rhodium-electrocatalyzed annulation study, the batch setup used a graphite felt anode and a platinum cathode. The flow configuration instead used a nickel plate cathode and porous graphite anode, incorporating a turbulence promoter to improve mass transport.
Commercial reactors may also restrict electrode material selection because of their mechanical design. The Ammonite 8, for example, pairs a 316 stainless steel cathode with a carbon-filled PVDF anode mounted on a copper backing plate. Other laboratory electrolyzer designs offer platinum electrodes as an anode option. Grooved electrode structures can also limit the use of brittle materials such as glassy carbon and boron-doped diamond.
Electrode structure can be just as important as material composition. In the FM01-LC laboratory electrolyzer, a flat-plate electrode produced a mass transfer coefficient close to 10⁻³ cm/s across the tested flow rates. A tenfold increase in flow rate produced approximately a threefold increase in limiting current density. Adding one or more thin polymer mesh turbulence promoters typically increased conversion rates by 1.5–3 times, with stacked meshes producing increases greater than fivefold in some experiments.
Performance varies across reactor designs. In the ElectroCell Microflow cell (MFC) and MP cell (MPC), turbulence promoters increased mass transfer by approximately 1.4 and 1.5 times, respectively. Woven mesh electrodes, however, increased reaction rates by 5.9–6.7 times, largely because of their greater surface area. The researchers suggested that fine woven meshes may be preferable for reactions limited by mass transport.
Three-dimensional electrodes, including foams, felts and mesh stacks, can provide substantially greater surface area within the reactor. In the FM01-LC, a 4 mm nickel foam cathode increased conversion rates by more than 20 times compared with a flat plate. However, increasing electrode thickness eventually produces diminishing returns. At higher current densities, ohmic losses within the electrolyte-filled structure can leave deeper regions underutilized, limiting productivity gains from additional thickness.
Recommended reading: Micro Solutions for Macro Challenges: Electroforming Applications in Energy, Medical, Food, and Beyond
Flow-Through vs. Flow-By Electrodes
Flow-through and flow-by configurations are two common approaches to distributing electrolyte in porous-electrode systems, particularly in redox flow battery research.
In a flow-through configuration, electrolyte passes through the porous electrode itself. This arrangement provides direct contact with the electrode structure and can improve reactant access, but thicker electrodes are often needed to maintain sufficient contact and surface area. Electrodes exceeding 3 mm are common in some designs, with compression managed to limit pressure drop. However, thicker structures can increase ohmic resistance and reduce energy efficiency at high current densities.
In a flow-by configuration, electrolyte moves through channels formed in or alongside the flow-field plate, with reactants reaching the porous electrode through shorter convection and diffusion pathways. This arrangement can support thinner electrodes and lower resistance.
In one redox flow battery study using carbon paper electrodes, a flow-by configuration with electrodes pressed directly against the flow field measured an area resistance of 0.5 Ω·cm², compared with 3.5 Ω·cm² for the flow-through configuration. These values reflect the specific cell designs and operating conditions tested, rather than a universal advantage of one architecture.
Inter-electrode Gap, Spacers, and Gaskets
The inter-electrode gap is a key design variable. In parallel-plate reactors, the spacer thickness determines the separation between the electrodes. Gaps of 1–5 mm are readily achievable, while some laboratory designs use considerably smaller distances. The Jud, Kappe and Cantillo cell uses 0.3 mm Mylar foils to define the channel, allowing the gap to be adjusted by changing the foil thickness. Narrower gaps can also support electrolysis with lower electrolyte concentrations, around 10 mM in some laboratory pipe-cell designs.
Reducing the gap, however, introduces practical limitations. Gas bubbles may accumulate in confined regions, interrupting ionic transport and creating localized resistance. Suspensions, precipitates and poorly soluble products can also obstruct narrow channels, making the reactor more susceptible to clogging.
Microfabricated screening platforms can use gaps as narrow as 10 µm between interdigitated platinum electrodes. Although these configurations offer compact reaction volumes and closely spaced surfaces, their narrow channels are vulnerable to clogging and generally require highly soluble reaction media.
Increasing the gap can reduce some flow restrictions, but it also increases ohmic losses. When gas evolution or solid formation makes narrow channels impractical, alternative designs such as recirculating reactors, rotating-cylinder reactors or gas-fed configurations may offer better operating conditions.
Membranes and Separators: Divided vs. Undivided Cells
An undivided cell allows both electrode reactions to occur within the same electrolyte compartment. A parallel-plate reactor can be converted into a divided cell by introducing a membrane or separator between the electrodes, along with the additional spacers, gaskets and compartment connections required to accommodate it.
A divided cell is useful when the products of one electrode reaction interfere with the reaction or products at the other electrode. Separation can prevent unwanted oxidation or reduction, limit product crossover and allow the anolyte and catholyte to be managed independently.
The commercial ElectraSyn Flow, for example, can be configured with a Nafion membrane for divided-cell electrolysis. In some mesh-electrode configurations, the electrodes are pressed against the membrane, while reactant is supplied through the back of the mesh. This arrangement is known as a zero-gap cell in the terminology used by Pletcher's group.
Zero-gap CO₂ electrolyzers commonly use anion-exchange membranes (AEMs), cation-exchange membranes (CEMs) or bipolar membranes (BPMs), depending on the desired ion transport and reaction environment.
Cation-exchange membranes (CEMs) contain negatively charged functional groups that repel anions while allowing cations to migrate through the membrane. Nafion, a widely used CEM, contains sulfonic acid groups.
Anion-exchange membranes (AEMs) contain positively charged functional groups that facilitate anion transport. Many commercial AEMs use quaternary ammonium groups.
Bipolar membranes (BPMs) combine an anion-exchange layer and a cation-exchange layer, enabling water dissociation at the interface and transport of hydrogen and hydroxide ions in opposite directions under appropriate conditions.
In CO₂ electrolysis, AEMs can support an alkaline environment at the cathode, which influences reaction selectivity and product formation. However, cation transport also affects cell performance. Cations crossing the membrane can contribute to carbonate formation, and solid carbonate deposits may accumulate in the pores of gas diffusion electrodes or flow fields, restricting reactant transport and reducing cell performance.
Membrane selection therefore involves trade-offs in ionic conductivity, selectivity, chemical stability, resistance, water transport and product crossover.
Recommended reading: PEM Fuel Cell: How It Works, Materials, and Design Trade-Off
Flow Fields and Manifolds
Flow fields and manifolds distribute electrolyte across the electrode surface while minimizing pressure drop and pumping requirements. Common designs include serpentine, interdigitated, parallel and spiral channels, with serpentine and interdigitated configurations among the most extensively studied.
Serpentine flow fields: The inlet and outlet are connected through a continuous channel. Pressure differences between adjacent channel segments drive some electrolyte through the porous electrode. Wider channels reduce pressure drop but may compromise mass transport.
Interdigitated flow fields: Inlet and outlet channels are separated, forcing electrolyte through the porous electrode between the channel ends. This arrangement promotes transport through the electrode but can increase pressure drop.
Flow field performance also depends on cell size and operating conditions. In a 400 cm² redox flow battery with an interdigitated design, a pressure drop of 0.05 MPa was reported at a flow rate of 2 mL/min per cm². More than 40% of the pressure drop originated from flow resistance between the inlet pipe and the ends of the branch channels. Narrow channels, such as those around 1 mm wide, and more viscous electrolytes can further increase pressure drop and pumping requirements.
Gas-fed electrochemical cells require a flow field on the gas side as well. In one CO₂ electrolysis design, a titanium flow field with a meandering channel distributed gas and provided electrical contact to the gas diffusion electrode (GDE).
Recommended reading: A microfluidic chip monitors gases using integrated, motionless pumps
Reference Electrode Selection and Placement
A reference electrode provides a stable potential against which the working electrode potential is measured. Its potential should remain sufficiently stable during operation and resist polarization when small currents pass through it. Selection depends on the electrolyte, solvent, temperature and whether the reference electrode introduces ions that could interfere with the reaction.
Reference electrode | Potential vs. SHE at 25 °C | Typical use and limitations |
Ag/AgCl, saturated KCl | 0.197 V | Widely used in aqueous systems; generally suitable up to approximately 80–100 °C, depending on configuration. Avoid where chloride contamination is undesirable. |
Saturated calomel (SCE) | 0.241 V | Traditional aqueous reference; temperature limitations apply, with Hg₂Cl₂ instability cited above 50 °C. |
Hg/Hg₂SO₄ in 0.5 M H₂SO₄ | 0.680 V | Useful when chloride ions must be avoided. |
Hg/HgO in 1 M NaOH | 0.140 V | Commonly used in alkaline electrolytes. |
Ag/AgNO₃, 0.1 M in acetonitrile | Approximately +0.36 V | Used in nonaqueous electrochemistry; potentials are often referenced internally to ferrocene. |
Reference potentials are dependent on electrode composition, electrolyte concentration and temperature. These values are representative, not universal.
Ag/AgCl electrodes are also available with 1 M or 3 M KCl filling solutions, which produce different reference potentials. Experimental reports should specify the filling solution and, where relevant, the temperature.
Narrow inter-electrode gaps limit the space available for a reference electrode. A Haber-Luggin capillary allows the reference electrode to be positioned close to the working electrode, reducing uncompensated resistance. However, positioning the capillary tip too close can obstruct flow or shield the working electrode, affecting local mass transport and current distribution. In the CO₂ flow cell discussed earlier, an Ag/AgCl reference electrode was introduced through a Haber-Luggin capillary in the catholyte supply line.
Capillary design also matters. Excessively narrow tips and high-impedance frits or salt bridges can increase resistance between the reference electrode and the cell. Frits and salt bridges can produce reference impedances of 20 kΩ or more, compared with approximately 1 kΩ for a typical Metrohm Ag/AgCl electrode. Such impedance can compromise potentiostat stability; Metrohm, for example, reports instability in High Speed mode above 30 kΩ on its Autolab systems.
Counter Electrode Selection
Platinum is commonly used as a counter electrode because it supports many electrode reactions with relatively low polarization and good chemical stability under suitable conditions. Carbon, copper and stainless steel may also be suitable, depending on the electrolyte and reaction chemistry. Counter electrodes for glass-cell experiments are available as platinum wire, coiled wire, mesh or graphite rods. In parallel-plate flow cells, the counter electrode is typically the second plate facing the working electrode.
Counter electrode geometry influences current distribution. A counter electrode with a shape and active area comparable to the working electrode can help distribute current more uniformly. For example, rectangular electrodes of similar dimensions or symmetrically arranged wire electrodes may be appropriate, depending on the cell configuration.
In a CO₂ electrolysis cell using a gas diffusion cathode, an Ir-MMO-coated titanium mesh served as a zero-gap counter electrode.
Counter electrode reaction products may interfere with the reaction of interest. In such cases, separating the electrode compartments can prevent unwanted interactions. A fritted tube can provide separation in conventional cells, although it may introduce resistance or other experimental limitations. In flow cells, a membrane or separator can provide compartmentalization while permitting selected ionic transport.
Materials of Construction
Materials in contact with the electrolyte, solvent, and reaction products must withstand the chemical and operating conditions of the cell. Commercial and laboratory flow cells use materials such as PEEK, PFA, ETFE, PTFE, Mylar, and chemically compatible elastomers, depending on the application.
For example, a spectroelectrochemical flow cell attachment uses PEEK and PFA/ETFE components, with EPDM O-rings for aqueous electrolytes and FFKM O-rings for organic solvents. Mylar is used in laser-cut cell assemblies because of its chemical resistance and suitability for constructing spacers and insulating layers. PTFE is also used in reactor compartments, including a modified anode compartment in an ElectraSyn Flow configuration.
Temperature and pressure ratings must be checked against the exact cell configuration. The Vapourtec Ion, when integrated with an E-Series or R-Series system, can operate between −10°C and 100°C and at pressures up to 5 bar. The stand-alone configuration does not include the same temperature-control components. In gas-fed electrochemical systems, pressure may also be regulated using membrane back-pressure regulators downstream of the cell, as demonstrated in the CO₂ flow cell discussed earlier.
Electrochemical Flow Cell Performance Metrics
Flow cells should be compared using measurable performance data, not descriptions alone. A 2024 review found that published reactor descriptions often omit important parameters, including interelectrode distance and effective electrode area. These details are necessary to assess performance and reproduce experimental results.
Metric | What it measures | Reported values and examples |
Current density (mA/cm²) | Electrical current per unit electrode area; an indicator of reaction rate | Aqueous-fed CO₂ cells typically operate below 60 mA/cm², with one mass-transfer estimate near 30 mA/cm². Commercial CO₂ electrolysis targets are often 200–500 mA/cm². A GDE flow cell reached 1.2 A/cm² using 2 M electrolyte, while a lithium-mediated ammonia cell operated at 6 mA/cm². |
Faradaic efficiency (%) | Fraction of total charge that produces the target product | CO₂ electrolysis often requires efficiencies above 95% for commercialization. A microfluidic GDE cell achieved 89% Faradaic efficiency for formic acid production. Some flow electrosynthesis reactions have achieved near-quantitative efficiency. |
Energy efficiency (%) | Fraction of electrical energy input converted into the chemical energy of the product | Reported values include 45% for a formic acid cell and 13% for continuous-flow lithium-mediated ammonia synthesis. |
Mass transfer coefficient, (k_m) (cm/s) | Rate at which reactants are transported to the electrode surface | Approximately 10⁻³ cm/s for a flat-plate electrode in the FM01-LC cell. Turbulence promoters increased conversion rates by 1.5–3 times, while a 4 mm nickel foam electrode achieved more than 20 times the conversion rate in the reported comparison. |
Area-specific resistance (Ω·cm²) | Resistance normalized to electrode area, reflecting ohmic losses across the cell | A redox flow battery study using carbon paper electrodes reported 0.5 Ω·cm² for a flow-by configuration and 3.5 Ω·cm² for a flow-through configuration. |
Pressure drop (MPa) | Pressure loss across the flow path, which affects pumping requirements | A 400 cm² interdigitated redox flow battery cell had a reported pressure drop of 0.05 MPa at 2 mL/min per cm². |
Conversion per pass (%) | Fraction of reactant converted during one transit through the cell | The Syrris microflow cell achieved more than 95% conversion at 1 mL/min and approximately 80% at 3 mL/min. |
Productivity | Quantity of product generated per unit time | The Syrris microflow cell produced 2 g/h at 3 mL/min. A small electrophotochemical flow reactor reported 0.41 mmol/h. |
Residence time (s) | Time a fluid element spends inside the reactor | Below 1 second per pass in some recirculating pipe cells, where conversion during each individual pass is correspondingly low. |
These values are examples from different reactor designs and reaction systems, not universal performance targets. The most relevant metric depends on the intended application and production scale. Faradaic efficiency may receive less attention in laboratory synthesis when product selectivity and isolated yield are the primary objectives. Energy efficiency becomes more consequential in large-scale production, particularly for high-volume, low-value chemicals.
Applications of Electrochemical Flow Cells
Flow Electrosynthesis
Electrosynthesis is attracting growing interest in organic chemistry, where reaction performance depends heavily on electrochemical cell design. Flow cells are suited to what Pletcher's group describes as convenient laboratory synthesis: producing approximately 10 mg to 100 g of product with high selectivity, ideally within minutes and generally within a few hours.
Several commercial flow reactors were available by 2024, although no single model had achieved the broad adoption of IKA's ElectraSyn 2.0 in batch electrochemistry. The following examples illustrate differences in configuration and operating capabilities.
Commercial flow reactor | Reported features |
IKA ElectraSyn Flow | Rectangular electrodes that do not require drilled fluid connections; can be configured with a Nafion membrane. |
Vapourtec Ion | Operates from −10 °C to 100 °C and up to 5 bar when integrated with a Vapourtec E-Series or R-Series system. The stand-alone configuration does not include the same temperature-control components. |
Syrris Asia | Operating range of 0–60 °C and pressures up to 5 bar; assembled without special tools and includes a built-in power supply. |
Ammonite 8 and Ammonite 15 (Cambridge Reactor Design) | Channel lengths of 1 m and 2 m, respectively, with a typical interelectrode gap of 0.5 mm. The Ammonite 15 achieved nearly a tenfold productivity increase in a reported deprotection reaction. |
ElectroCell Microflow Cell (MFC) and MP Cell (MPC) | Smaller models in a range of commercial flow reactors designed to support scale-up. |
Many research groups still build their own flow cells. Low-cost fabrication methods include 3D printing and laser cutting. One reported 3D-printed reactor cost less than US$5, while laser-cut components offer an alternative to more expensive CNC machining.
For larger production volumes, numbering-up provides another scale-up strategy. Rather than enlarging a single reactor, multiple smaller reactors operate in parallel. The Xu group used 20 undivided parallel-plate reactors to scale an electrochemical aromatic hydroxylation reaction to 250 g (1.44 mol) of product.
CO₂ Electrolysis and Gas-Fed Flow Cells
CO₂ reduction illustrates how reactor design can limit electrochemical performance. Carbon dioxide has low solubility in water, approximately 34–35 mM under ambient conditions. In aqueous-fed cells, limited dissolved CO₂ and slow transport to the catalyst constrain current density, which is typically below 60 mA/cm².
A gas diffusion electrode (GDE) addresses this limitation by supplying gaseous CO₂ directly to the catalyst. This shortens the transport distance compared with diffusion from dissolved CO₂ in the bulk electrolyte. One reported comparison describes a diffusion path of approximately 50 nm from the gas side to the catalyst, compared with about 50 µm from the bulk electrolyte.
A GDE typically combines a porous gas diffusion layer with a catalyst layer. Planar designs may contain three functional layers: a superhydrophobic macroporous layer, a microporous layer and the catalyst layer. PTFE can also be incorporated into the catalyst layer. In one copper–PTFE electrocatalyst, it helped limit water penetration while preserving pathways for CO₂ transport to the reaction sites.
Two common gas-fed configurations are used in CO₂ electrolysis:
Membrane electrode assembly (MEA): The cathode GDE contacts the membrane directly, reducing the distance between the catalyst and the membrane.
Flowing-catholyte configuration: The GDE contacts a liquid electrolyte that flows through the cathode compartment.
Rotating disc electrodes and H-cells remain useful for controlled catalyst studies, but continuous-flow electrolyzers are needed to evaluate sustained operation and develop scalable CO₂ conversion processes.
One persistent challenge is GDE flooding, which occurs when liquid electrolyte penetrates the porous electrode and gas channel. Flooding can restrict gas transport and interrupt operation. In a study by Martens and co-workers, a GDE remained resistant to flooding throughout the reported tests, allowing current densities of up to 1.2 A/cm² with a 2 M electrolyte. Under those experimental conditions, the GDE itself did not appear to limit performance. Electrolyte heating and gas evolution in the electrolyte gap instead constrained the maximum current density.
Recommended reading: What Is a Fuel Cell? Theory, Components, Types, and Applications
Electrochemical Detection in Flow (LC-EC)
Liquid chromatography with electrochemical detection (LC-EC) uses thin-layer flow cells in which the sample passes through a narrow space between the working and auxiliary electrodes. The gasket separating the electrodes determines the cell volume and liquid-layer thickness. BASi, for example, supplies PET and PTFE gaskets ranging from 13 µm to approximately 381 µm, including 13, 51 and 127 µm options. Gaskets can be used individually or stacked to adjust the cell thickness.
Another detector configuration is the wall-jet electrode, in which a liquid stream strikes a planar electrode surface perpendicularly and spreads radially. This geometry is used in flow injection analysis and HPLC detection. Metrohm's ion chromatography amperometric detector, for example, offers wall-jet cell kits with gold working electrodes for carbohydrate analysis and silver working electrodes for cyanide detection. Both configurations use a palladium reference electrode and support several measurement modes, including DC, pulsed amperometric detection (PAD), flexible integrated pulsed amperometric detection (flexIPAD) and cyclic voltammetry (CV).
Spectroelectrochemistry and Operando X-Ray Cells
Flow cells can combine electrochemical measurements with optical or X-ray analysis to monitor reactions during operation. Redox.me's Spectro-EFC attachment, for example, couples screen-printed electrodes to a UV-vis-NIR light source and spectrometer through an optical fiber. The system is designed for in-situ detection of electrogenerated intermediates, including short-lived species. In flow mode, tubing connects the cell to a pump, typically a peristaltic pump; in static mode, electrolyte is introduced using a syringe.
Operando X-ray measurements require careful control of the electrolyte layer because water absorbs X-rays. Paparoni and co-workers reported in their preprint that a 1 mm water layer attenuates approximately 41% of a 10 keV photon flux, while only about 1% passes through at 5 keV. Their 2024 flow cell reduces the electrolyte thickness to approximately 17 µm. Continuous electrolyte movement also helps limit bubble formation and large reactant concentration gradients during measurement.
Dissolution Testing: Scanning Flow Cell ICP-MS
A scanning flow cell (SFC) coupled to inductively coupled plasma mass spectrometry (ICP-MS) measures metal dissolution during electrochemical operation. A metal sample, gradient alloy or catalyst ink deposited on a substrate is positioned beneath the small flow cell. A movable stage allows the operator to scan different sample locations. As the applied potential or current drives dissolution, the flowing electrolyte carries dissolved metal downstream to the ICP-MS, which measures the resulting signal over time.
The small contact area supports rapid measurements across different materials or surface compositions. In a platinum dissolution study by Topalov and co-workers, a V-shaped polyacrylate flow channel contacted approximately 1.1 mm² of the electrode surface. The delay between dissolution at the electrode and detection by the ICP-MS was approximately 15–20 seconds.
The same study used cyclic voltammetry in 0.1 M HClO₄, with the lower potential limit held at 0.05 V versus the reversible hydrogen electrode (RHE). Platinum dissolution became detectable when the upper potential limit exceeded approximately 1.1 V versus RHE. The researchers also found that significant dissolution was associated with transitions between oxidized and reduced platinum surface states across their room-temperature experiments, with upper potential limits reaching 1.55 V versus RHE.
Energy Storage: Redox Flow Cells
Each cell in a redox flow battery is an electrochemical flow cell. A typical design combines graphite plates, porous electrodes, and an ion-conducting membrane or separator. Research on redox flow batteries has also informed the understanding of mass transport, electrolyte distribution, and concentration gradients in parallel-plate electrochemical reactors.
Single-cell testing is used to evaluate battery materials and components before stack development. For example, Pacific Northwest National Laboratory (PNNL) evaluated electrolyte, membrane or separator, carbon felt electrodes, and bipolar plates in single-cell configurations as part of development toward kilowatt-scale stacks.
Recommended reading: Flow Cell Technology: A Comprehensive Guide
Scaling Up: From Laboratory Flow Cell to Stack
Laboratory electrochemical flow cells typically use a single pair of electrodes, with active areas reaching approximately 100 cm². Industrial parallel-plate electrolyzers can use electrodes with areas of up to 1 m², while filter-press assemblies may stack as many as 100 cells using bipolar electrical connections.
Moving from laboratory experiments to larger production volumes generally involves two scale-up strategies: numbering up and sizing up.
Numbering up: Multiple reactors or electrode pairs operate in parallel to increase total production capacity. This approach retains smaller individual reactor units but requires consistent flow distribution across them. Uneven distribution can produce differences in residence time, mass transfer and conversion, particularly in multiphase systems.
Sizing up: The dimensions or active electrode area of an individual reactor increase. Extending the flow channel can preserve its cross-sectional geometry and surface-to-volume ratio but increases channel length and pressure drop. Increasing reactor width or electrode area also requires attention to current distribution, electrolyte flow and heat transfer.
Scaling up also requires integration of the reactor with pumps, piping, valves, flow controls, power supplies and monitoring equipment. These connections must be designed to support stable operating conditions, maintain uniform electrolyte distribution and accommodate the thermal and hydraulic demands of a larger system.
Commercial reactor families provide options for evaluating larger configurations. For example, ElectroCell offers flow reactors in different sizes to support process development and scale-up. The Ammonite 15 has also shown higher productivity in reported electrochemical synthesis applications than the smaller reactor configuration, though the improvement depends on the reaction and operating conditions.
How to Specify an Electrochemical Flow Cell
Use the following steps to translate reaction requirements into a practical flow cell specification.
Define the production target. Specify the required product rate, conversion and selectivity. These targets determine the approximate electrode area, cell dimensions and operating conditions.
Calculate the required current and electrode area. Apply Faraday's law using the reactant concentration and volumetric flow rate, as shown in the worked example above. Divide the required current by the sustainable current density to estimate the necessary active electrode area. Account for Faradaic efficiency when calculating the actual current requirement.
Select the operating mode. Single-pass operation suits continuous production and integration with downstream processes. Recirculation allows repeated passes through the reactor and can support higher circulation rates for bubble removal or reactions where lower conversion per pass is acceptable.
Choose the interelectrode gap. Start with the narrowest practical gap that maintains reliable electrolyte flow. Laboratory spacers commonly provide gaps of 1–5 mm, while some flow electrolysis cells operate below 1 mm. Allow sufficient space for gas bubbles, suspended solids, and precipitates where these are expected.
Select electrode material and structure. Flat plates provide a well-defined geometry and current distribution. Meshes, foams and felts offer greater surface area and can improve mass transfer, although thick porous electrodes may introduce additional ohmic losses.
Decide whether the cell should be divided or undivided. Use a membrane or separator when counter electrode products could interfere with the desired reaction. Select the membrane according to the required ion transport, chemical compatibility, and operating conditions.
Choose the flow field and manifolds. Evaluate how evenly the electrolyte reaches the active area, how effectively it penetrates porous electrodes, and the pressure drop required to maintain the target flow rate.
Specify electrochemical control. Select a reference electrode compatible with the electrolyte and temperature, and plan its position and connection to minimize uncompensated resistance. Choose a potentiostat for controlled electrode potential or a suitable constant-current power supply for galvanostatic operation.
Check materials and operating limits. Confirm that the cell body, seals, spacers, gaskets, and tubing are compatible with the electrolyte and solvent. Verify the temperature and pressure ratings of the exact reactor configuration.
Document the reactor configuration. Record the interelectrode distance, effective electrode area, electrode materials, flow rate, electrolyte composition, operating mode, and temperature. These parameters are essential for interpreting results and reproducing experiments.
Common Mistakes and Troubleshooting
Problem | Why it occurs | What to check or change |
Gas blocking the electrode gap | Bubbles disrupt electrolyte flow, reduce active electrode area and create regions of high resistance. | Increase circulation where appropriate to remove bubbles. A rotating-cylinder electrode can also promote bubble detachment and reduce bubble accumulation. |
Channel clogging | Precipitates, suspended solids or poorly soluble reactants and products obstruct narrow flow paths. | Check solubility throughout the reaction, review electrolyte compatibility and avoid channel dimensions that cannot accommodate expected solids. Design the cell for cleaning and maintenance. |
Electrolyte leaks | Insufficient or uneven compression, damaged gaskets or poorly sealed fittings allow liquid to escape. | Apply uniform compression across the assembly, inspect gasket alignment and condition, and check fittings before operation. |
Stray current paths | Inadequate electrical insulation allows current to pass through conductive end plates or other unintended components. | Use suitable insulating layers between the electrodes and metal end plates, and verify electrical isolation before operation. |
Potentiostat instability or oscillation | High reference-electrode impedance, poor electrical connections or an unsuitable measurement configuration can destabilize potential control. | Check the reference compartment and electrical connections. Avoid excessively narrow Luggin capillary tips and unnecessary high-impedance frits. For Metrohm Autolab instruments, follow the manufacturer's recommendations for reference impedance and High Stability settings. |
GDE flooding and salt deposition | Liquid penetration blocks gas pathways, while carbonate precipitation can obstruct GDE pores and flow channels during CO₂ electrolysis. | Monitor electrolyte pressure and water management, assess membrane and GDE compatibility, and adjust operating conditions to limit flooding and salt accumulation. |
Poor utilization of thick porous electrodes | Ohmic losses can prevent deeper regions of a foam or felt from contributing effectively at high current densities. | Optimize electrode thickness, compression and conductivity. Compare performance against thinner structures at the same operating conditions. |
Electrode degradation | Corrosion, dissolution, passivation or mechanical deterioration can reduce performance during prolonged operation. | Monitor electrode condition and performance over time, and evaluate material stability under the intended electrolyte, potential and temperature conditions. |
Uneven flow distribution | Poor inlet and outlet geometry creates preferential flow paths and leaves parts of the electrode underutilized. | Include appropriate entry and exit lengths or purpose-designed flow distributors to improve electrolyte distribution across the active area. |
"Flow Cell" in Other Fields: Sequencing, Optical and Cytometry Flow Cells
The term flow cell refers to different devices across scientific fields. In DNA sequencing, optical spectroscopy and flow cytometry, it describes components with functions and design requirements distinct from those of electrochemical reactors.
In Illumina sequencing, the NextSeq 550 flow cell is a glass-based substrate housed in a cartridge. DNA clusters form on its surface, where sequencing reactions take place. Each cluster is a clonal group of template DNA molecules attached to the flow cell surface.
The NextSeq 550 flow cell has four lanes, imaged in pairs, with each lane divided into smaller imaging areas called tiles. MiSeq System flow cells have a single lane, while the number of tiles varies by flow cell type.
Oxford Nanopore sequencing uses a different design. Its flow cell is a consumable inserted into the sequencing device, containing thousands of nanopores embedded in an electrically resistant membrane. DNA or RNA passes through individual pores, and the sequence is determined from characteristic changes in ionic current.
In optical spectroscopy, a flow cell is a flow-through cuvette that allows a liquid sample to pass through an optical measurement path. Hellma, for example, offers quartz flow-through cells with optical path lengths ranging from 0.1 mm to 50 mm. Its product range also includes a quartz flow-through cuvette with a superfine channel for flow cytometry.
Although these devices share the term flow cell, their operating principles and design requirements differ from electrochemical flow reactors. For sequencing, spectroscopy, and cytometry applications, consult the relevant instrument manufacturer's documentation.
Recommended reading: Solid Oxide Fuel Cell: How It Works, Materials, and Design Trade-Offs
Conclusion
An electrochemical flow cell uses controlled electrolyte flow and closely spaced electrodes to improve reaction conditions, mass transport and electrode utilization. Its performance depends on balancing several design variables: electrode spacing against gas and solids management, porous electrode area against internal ohmic losses, and flow-field performance against pressure drop.
Begin with the reaction requirements. Calculate the current using Faraday's law, account for Faradaic efficiency, and check whether the electrode structure can sustain the required current density through adequate mass transfer. Then select the operating mode, electrode configuration, membrane, reference electrode and flow field to suit the electrolyte and reaction.
Finally, document the interelectrode gap, effective electrode area, electrode materials, electrolyte composition and flow rate. These details make performance comparisons meaningful and allow other researchers to reproduce the reactor configuration, whether it is used for electrosynthesis, CO₂ reduction, electrochemical detection or dissolution analysis.
Frequently Asked Questions
What is a flow cell?
In electrochemistry, a flow cell is an electrochemical reactor in which a pump moves electrolyte through a channel between electrodes. Flow electrolysis cells often use narrow interelectrode gaps, with some designs operating below 1 mm. The term also describes devices in other fields, including Illumina's glass-based NextSeq sequencing flow cells and optical flow-through cuvettes.
How does an electrochemical flow cell work?
A pump moves electrolyte from a reservoir through the cell, where electrochemical reactions occur at the working and counter electrodes. A potentiostat or power supply controls the electrode potential or current, depending on the operating mode. The electrolyte either exits as a product stream in single-pass operation or returns to the reservoir for recirculation. Closely spaced electrodes and controlled flow can improve mass transport and reduce ohmic losses, depending on the cell design and operating conditions.
What is the difference between a flow cell and an H-cell?
An H-cell takes its name from its shape. It typically has separate anode and cathode compartments connected through an ion-conducting membrane or separator. The electrolyte is generally stationary during operation, unlike in a flow cell, where a pump moves it through a defined channel. H-cells are useful for laboratory experiments that require compartmentalization, while flow cells offer greater control over electrolyte movement and support continuous processing. Both designs have limitations that depend on their geometry, electrode arrangement and operating conditions.
Which reference electrode should I choose?
For many aqueous electrochemical experiments, Ag/AgCl with saturated KCl is a common choice. Its reference potential is approximately 0.197 V versus the standard hydrogen electrode (SHE) at 25 °C. Hg/HgO is commonly used in alkaline electrolytes, while Hg/Hg₂SO₄ is an option when chloride contamination must be avoided. For nonaqueous solvents, Ag/AgNO₃ can be used, often with ferrocene as an internal reference. Always report the reference electrode composition, filling solution and temperature because these affect the measured potential.
What counter electrode do I need?
Platinum is a common counter electrode material because it supports many electrode reactions and is chemically stable under suitable conditions. Carbon, copper and stainless steel may also be appropriate, depending on the electrolyte and reaction chemistry. Choose a geometry and active area that provide adequate current distribution. If counter-electrode products could interfere with the reaction of interest, separate the compartments using a fritted tube in a conventional cell or a membrane or separator in a flow cell.
Are electrochemical cells electrolytic?
Not all electrochemical cells are electrolytic. A galvanic (voltaic) cell converts energy from a spontaneous chemical reaction into electrical energy. An electrolytic cell uses an external electrical source to drive a nonspontaneous reaction. A rechargeable battery operates in both modes: it behaves as a galvanic cell during discharge and undergoes an electrically driven reverse reaction during charging. A redox flow battery therefore operates galvanically during discharge and electrolytically during charging.
Which way do electrons flow in an electrochemical cell?
Oxidation occurs at the anode and reduction at the cathode. In a galvanic cell, electrons flow from the anode to the cathode through the external circuit. In an electrolytic cell, an external power supply drives the reaction: it draws electrons from the anode and supplies electrons to the cathode. The electrode names remain tied to the reactions, but their electrical signs differ between the two cell types.
Recommended reading: Cathode Anode: The Basics and Applications
What is a redox flow cell?
A redox flow cell is the electrochemical unit within a redox flow battery. A typical design uses graphite plates, porous electrodes and an ion-conducting membrane or separator. Multiple cells are connected in a stack to increase battery voltage and capacity.
What interelectrode gap should a flow cell have?
Use the narrowest gap that your reaction and operating conditions can accommodate. Parallel-plate cells can readily be assembled with gaps of 1–5 mm, while some flow electrolysis designs operate below 1 mm. Reported examples include an extended-channel cell with a gap of approximately 200 µm and Ammonite cells with a 0.5 mm gap.
Gas bubbles, suspended solids and precipitates can make very narrow channels difficult to operate. Increasing the gap can reduce flow restrictions but also increase ohmic losses. Where gas or solids are difficult to manage, consider alternatives such as recirculation, a rotating-cylinder reactor or a gas-fed cell.
References
Regnier, M., Vega, C., Ioannou, D. I., & Noël, T. (2024). Enhancing electrochemical reactions in organic synthesis: The impact of flow chemistry. Chemical Society Reviews, 53(21), 10741–10760. https://doi.org/10.1039/D4CS00539B
Pletcher, D., Green, R. A., & Brown, R. C. D. (2018). Flow electrolysis cells for the synthetic organic chemistry laboratory. Chemical Reviews, 118(9), 4573–4591. https://doi.org/10.1021/acs.chemrev.7b00360
Sun, J., Guo, Z., Pan, L., et al. (2023). Redox flow batteries and their stack-scale flow fields. Carbon Neutrality, 2, Article 30. https://doi.org/10.1007/s43979-023-00072-6
Garg, S., Giron Rodriguez, C. A., Rufford, T. E., Varcoe, J. R., & Seger, B. (2022). How membrane characteristics influence the performance of CO₂ and CO electrolysis. Energy & Environmental Science, 15(11), 4440–4469. https://doi.org/10.1039/D2EE01818G
Martens, C., Schmid, B., Tempel, H., & Eichel, R.-A. (2023). CO₂ flow electrolysis: Limiting impact of heat and gas evolution in the electrolyte gap on current density. Green Chemistry, 25(19), 7794–7806. https://doi.org/10.1039/D3GC02140H
Paparoni, F., Alizon, G., Zitolo, A., Rezvani, S. J., Di Cicco, A., Magnan, H., & Fonda, E. (2024). A novel electrochemical flow-cell for operando XAS investigations in X-ray opaque supports. Physical Chemistry Chemical Physics, 26(5), 3897–3906. https://doi.org/10.1039/D3CP04701F
Flowers, P., Theopold, K., Langley, R., & Robinson, W. R. (2019). Chemistry 2e. OpenStax. Section 17.7, Electrolysis. https://openstax.org/books/chemistry-2e/pages/17-7-electrolysis
in this article
1. Key Takeaways2. Introduction3. What Is a Flow Cell in Electrochemistry?4. How an Electrochemical Flow Cell Works5. Electrochemical Flow Cell Design: Components and Architecture6. Electrochemical Flow Cell Performance Metrics7. Applications of Electrochemical Flow Cells8. How to Specify an Electrochemical Flow Cell9. Common Mistakes and Troubleshooting10. "Flow Cell" in Other Fields: Sequencing, Optical and Cytometry Flow Cells11. Conclusion12. Frequently Asked Questions13. References